Which sucks more? Comparison of two novel direct-in-scope suction ureteroscopes......
发布时间:2025-08-20 阅读次数:53

E0B6E6C4-D9D4-4eb6-AF68-2F6C2EB989FC 拷贝.png


Authors: Zhamshid Okhunov, Elizabeth A Baldwin, Daniel Jhang, Ruben Crew, Ninous Betdashtoo, Kallan Richards, Grant Sajdak, Joshua Ghoulian, D Duane Baldwin 


Abstract


Introduction: Novel Direct-ln-Scope-Suction ureteroscopes (DlSS-U) have recently been introduced to facilitate removal of small stone fragments. The purpose of this study was to compare two novel DlSS-U to a conventional ureteroscope to determine efficiency for suctioning fluid and stone dust at different deflection angles.


Methods: Three disposable ureteroscopes were evaluated, including two DISS-U with 5.1-Fr and a 3.6-Fr channels, and  were compared to a conventional ureteroscope with 3.6-Fr channel. For DISS-U trials, a conventional suction system was  set at maximal settings and tested at deflection of 0, 90, and 140°, to simulate upper, interpolar, and lower pole kidney stone  evacuation under 3 conditions (empty channel, 2.2-Fr basket, and 0.038 guidewire). For the conventional ureteroscope, suctioning was performed manually using a 60-mL Luer lock syringe. Five trials were performed to measure the time in seconds  to remove 1 gram of CaOx density BegoStones (0.25, 0.5, and 1 mm) and 50-mL of saline.


Results: The 5.1-Fr DISS-U was the only scope that evacuated 0.5 mm stone fragments and evacuated them significantly  faster from the upper and interpole compared to the lower pole (111.0±27.7, 128.0±29.1, and 377.7±63.4 s at 0, 90, and  140°, respectively; p<0.002). The 5.1-Fr DISS-U evacuated 1 g of dust faster than the 3.6 DISS-U and conventional ureteroscope at all angles (p<0.007). The 5.1-Fr DISS-U and the 3.6-Fr DISS-U required less suction time for 50 mL of saline  compared to the conventional ureteroscope at all angles (p<0.001 for all). Using a 2.2-Fr basket increased suction time by  66.0% (p<0.001) for the 3.6-Fr DISS-U and by 44% (p<0.001) for the 5.1-Fr DISS-U, while the 0.038 guidewire increased  suction time by 96% and 111%, respectively (p<0.001). No ureteroscope was able to suction 1 mm stone fragments without  clogging.


Keywords: Direct in scope suction; Lithotripsy; Ureteroscopy; Urolithiasis.



阅读原文 >>>



产品中心
  • 外科产品线
  • 内科产品线
  • 配件
新闻中心
  • 会议要闻
  • 企业动态
学术前沿
  • 一次性输尿管软镜
  • 文献分享
走进普生
  • 发展历程
  • 行销全球
  • 公司荣誉
  • 大爱普生
联系我们
  • 联系我们
  • 招贤纳才
  • 我在普生

扫一扫,关注我们

了解更多最新资讯

广东省珠海市高新区科技三路33号3栋5楼 

400-688-2023  

pusen@pusenmedical.com

©2025 珠海普生医疗科技股份有限公司  All Rights Reserved  
 粤网药信备
字〔2025〕第00816号  1710212326908180.png粤公网安备44049102496975号   

反馈收集单